A comparative study of the efficacy and comfort of the Purewick™ male external catheter in healthy volunteers

Wagg A1, Sena M2

Research Type

Clinical

Abstract Category

Continence Care Products / Devices / Technologies

Abstract 645
Open Discussion ePosters
Scientific Open Discussion Session 105
Thursday 24th October 2024
14:25 - 14:30 (ePoster Station 5)
Exhibition Hall
New Devices Conservative Treatment Male Prospective Study
1. Department of Medicine, University of Alberta, 2. Becton-Dickinson & Co.
Presenter
Links

Poster

Abstract

Hypothesis / aims of study
Prolonged use of indwelling catheters is associated with urinary tract infection and urethral trauma in men.  The use of an external urine collection device may avoid these complications.  For males, condom catheters are an option for external drainage, however, there are  limitations to their use.  The increasing prevalence of men with a buried or hidden penis secondary to obesity or, in the acute hospital setting, with generalized oedema is an increasing problem. Obesity has a prevalence of up to 40.8% in middle-aged men in the United States and, in one series, 87% of men who went surgical management for their acquired buried penis was obese [1,2]. 
An alternative device, the PureWick™ male external catheter is intended for non-invasive urine output management in male patients may be suitable for use in obese men. There are no data on the performance of this device in men unable to use conventional condom catheters. This study sought to assess the efficiency of urine collection over two voids in a group of obese and non obese men.
Study design, materials and methods
In this prospective, post-market, crossover, single-blind, single center study, healthy male volunteers were randomized 1:1 to a treatment sequence using two devices (PureWick™ Male External Catheter and Sage PrimoFit™) and were followed for 1 day through 2 voids. Performance was defined a priori as the percentage capture of urine following a voluntary void (amount of urine captured by device and collected in canister, measured by weight expressed as a percentage of the total void, captured urine and leaked urine weight, measured by change in absorbent pad weight).
Secondary aims comprised assessment of the performance of the PureWick™ system in men unable to use traditional sheath style condom catheters, participant comfort after using the device and health care providers’ perspectives on ease of use in placing and removing the device, both measured by 5 point Likert scale. 
To be eligible for inclusion men had to be 18 years of age or older, to have the ability to speak and understand English, to adhere to the required study procedures and to independently void urine. Men with urinary incontinence which prevented spontaneous voiding; urinary retention, frequent episodes of bowel incontinence, any irritation, wound, open lesion, at the device application site, or on the genitalia,  an inability to comply with study procedures independently were excluded. Approximately 50% of the men recruited were obese so that potential differences based on anatomy could be examined. Men were divided into non-morbidly obese men (BMI <40 kg/m2) or morbidly obese men (BMI ≥40 kg/m2). 
Participants were randomized 1:1 to one of the two treatment sequences. Randomization was stratified by non-morbidly obese men (BMI <40 kg/m2) or morbidly obese men (BMI ≥40 kg/m2). 
 A sample size of 44 men provided 90% power to detect a mean paired difference in capture rate of half of the standard deviation of the paired difference in capture rate of 5% using a two-sided paired t-test at significance level of 0.05.
The investigational devices are commercially available Class I, 510(k) exempt devices intended for non-invasive urine output management in male patients. 
Prior to arrival, participants were instructed to consume fluids and to come to the study site with a full bladder. Once the participant felt need to void, depending on randomized treatment order, the HCP placed either the PureWick™ MEC or the Sage PrimoFit™ device. Participants were monitored for approximately 2 hours from the time of the device application, through one void. After the void, the pre-weighed absorbent pad and urine collection canister were reweighed. The device was removed and the HCP noted any signs of irritation/injury in the perineal area.The participant was asked to complete a brief survey to assess device comfort immediately after the completed void. The HCP completed a brief survey to assess device ease of use for the participant. Following this process, participants were asked to drink and when the participant felt the need to void a second time, the HCP confirmed that the bladder volume was >125mL by portable bladder scanner. Void 2 was repeated using the 2nd device. Particiapnt and HCP surveys were repeated.
Results
Fifty nine men completed the study. The mean (standard deviation, SD) age of the men was 40.8 (SD12.4) years, with a median of 37.0 years. Other demographic variables are shown in Table 1. For void 1, the mean (SD) proportion of urine capture for the Purewick device was 97.8 (10.0) %, for the Primofit device this value was 90.7 (20.7)%.  The mean (95% CI) difference between devices was 6.6 (0.18, 13.0) %, p=0.044. For the second void, the mean (SD) percentage capture was 91.1 (25.8)% for the Purewick device and 85.2 (21.7)% for Primofit. A 100% capture rate was seen for 39/59 (66.1%) of Purewick voids and 15/59 (25.4%) Primofit voids. For the morbidly obese men, for the first void, these figures were 99.7 (0.88) %, (Purewick) versus 83.5 (32.2) % (Primofit) and for the second void 80.8 (36.9) % (Purewick) and 79.3 (23.3) % (Primofit). To assess whether voided volume (weight) affected performance, the difference in the difference in mean voided urine weight (g) between devices was examined but no different, p=0.24.   Given the generally lower voided weights in void 2, the difference in void weight between void 1 and 2 was assessed (mean (95%CI) difference, -63.8 (-127.7, 0.07)),  but not different, p=0.05. These relationships held for the morbidly obese group. The scores for the participant comfort Scale and professional’s ease of use scales are shown in Table 2. There were no adverse events reported in the study.
Interpretation of results
There was a statistically significantly superior performance between the capture rate of the investigational device (Purewick MEC) and the comparator device, regardless of voided volume (weight) order of void or BMI category.  The median capture rate in morbidly obese men was higher for the Purewick device for both voids.   Assessed by simple Likert Scale, healthcare professionals scored both devices as somewhat easy or easy to place in 56/59 (95%) of men.  They scored the Primofit device as easier to remove as compared to the Purewick device.   The men’s opinions of ease of use of the Purewick system are in line with a recent report on patient and caregiver satisfaction with use of the system [3]. This is the first study to assess the efficiency of urine capture using the male external device.  Strengths of this study include its controlled, comparative design with an adequately powered sample size but whilst intended to assess the utility of the Purewick device in men with buried penis, there were too few men with buried penis in the sample to make this assessment.  The level performance of the device in morbidly obese men is, though, reassuring.  This experimental study was a short-term assessment, under controlled conditions.  Longer term, pragmatic studies of the use of the male device in real life clinical applications are required which take into account daily use and shifting position.
Concluding message
This single blind cross over study of the efficiency of urine capture showed a statistically significantly superior performance of the Purewick device in males versus a commercially available comparator.  This relationship held regardless of voided urine volume and obesity status of the participating men.
Figure 1 Table 1. Demographics
Figure 2 Table 2. User and HCP ease of use
References
  1. Prevalence of Obesity and Severe Obesity Among Adults: United States, 2017–2018, in NCHS Data brief 2020, Centres for Disease Control and Prevention, National Center for Health Statistics.
  2. Curr Urol Rep, 2018. 19(3): p. 22.
  3. Can J Urol, 2022. 29(4): p. 11216-11223
Disclosures
Funding Trial funded by Becton-Dickinson Clinical Trial Yes Public Registry No RCT Yes Subjects Human Ethics Committee Advarra, Columbia MD, USA Helsinki Yes Informed Consent Yes
19/11/2024 19:40:17