Adjustable artificial sphincter VICTO in the treatment of urinary incontinence: early results in a cohort of high-risk cases in a prospective single center study

Ameli G1, Weibl P1, Rutkowski M1, Hübner W1

Research Type

Clinical

Abstract Category

Male Stress Urinary Incontinence (Post Prostatectomy Incontinence)

Abstract 256
Open Discussion ePosters
Scientific Open Discussion Session 18
Friday 9th September 2022
11:25 - 11:30 (ePoster Station 2)
Exhibition Hall
Male Stress Urinary Incontinence Incontinence Surgery
1. Department of Urology, Teaching Hospital, Clinic of Korneuburg, Austria
In-Person
Presenter
G

Ghazal Ameli

Links

Poster

Abstract

Hypothesis / aims of study
Victo is an artificial urinary sphincter (AUS), consisting of three parts which are preconnected. A pressure regulating balloon, a control pump, and a cuff. Victo+ is offering an additional, so called stress regulating balloon, in case of short term increase of abdominal pressure such as coughing, additional fluid is provided by the stress regulating balloon to the cuff in order to increase temporarily the system pressure. Additionally, both systems offer the possibility of postoperative adjustment of the device pressure.
The objective of the present study is to test the efficiency and practicality of VICTO and VICTO+ in the control of stress urinary incontinence and to evaluate the short- and long-term complications.
Study design, materials and methods
A total of 30 patients were included between January 2020 and December 2021, the trial is still ongoing and estimated end of the clinical trial is May 2025. 
43,3% of the patients in this cohort has undergone prior procedures for stress urinary incontinence (SUI): sling n=4, ProACT balloons n=4, AMS 800=1, 3 patients had multiple prior SUI surgeries. 40% of the patients had undergone secondary irradiation therapy after prostatectomy. Additionally, 9 patients had undergone procedures for bladder neck pathologies or urethral strictures (UTI, TW n=7, end-to-end anastomosis n=2). 
All patients underwent artificial urethral sphincter (AUS) implant “VICTO” according to established technique. 
The mean age was 71,4 years, 40% of the patients had systemic arterial hypertension and 24% diabetes mellitus.
Functional urinary outcomes were assessed according to daily pad use and a standardized questionnaire. Data were collected as part of the follow up care of these patients.
Results
The reduction in number of pads compared between baseline and Follow-Up was significant (p<0,001). Cure of urinary incontinence was defined as using either 0 or 1 pad/day. All the index patients (no prior surgery, no previous irradiation) in this cohort (n=5) were cured. In total 
62,1% (n=18) of the patients reported to be dry and were cured by definition. Additionally, 20,7% (n=6) patients reported improvement of ≥ 50 % according the PGI-I (Patients Global Impression of Improvement). 
Further surgery has been performed in a total of 8/29 (27,6%) patient. One patient required the change of the pump due to malfunction and one patient required relocation of the pump due to facilitate the handling. Explantation of the device has been performed in four patients, the reasons were persistence of incontinence (n=2; 6,8%) infection and erosion (n=2; 6,8%). One of the infections occurred after traumatic catheterization without deactivation of the device which caused urethral erosion and consequently infection and explantation of the device. Both patients with erosion had previous pelvic irradiation. 
One explantation was reported at day 2 after implantation, the presumable cause of explantation was unrecognized intra-operative urethral injury and consequently urethral erosion and was excluded from the study.
Interpretation of results
Regarding the baseline characteristics such as age, body mass index, coronary heart disease, diabetes mellitus, history of urethral disease, result from this trial are in line with literature. However, considering history of pelvic irradiation and history of prior surgical treatment of SUI, the incidence was 43,3% and 40,0% respectively, which is at and way even above the upper limit in comparison to literature. Importantly, taking to account the current evidence on the impact of pelvic irradiation on the AUS outcome, the patient population in the current trial might not be only at increased risk for repeated surgery and explantation but also for increased risk of incontinence persistence, thus, decreased treatment success. 
The most common complication rates in artificial urinary sphincters are erosion, malfunction, dislocation and subsequently explantation. The overall combined erosion and infection rates were 6,8% in the current trial which are consistent with literature. Furthermore, prolonged catheterization or history of pelvic irradiation or urethral stricture are common risk factors for the development of urethral erosion.
Overall explantation rate in the current trial was 13,8%. It should be considered, that the explantation rates are consistent with literature even considering the increased risk for failure of the current population.
Overall treatment success was 82,8% according PGI-I and 62,1% according to pad per day usage.  In this complex group with mostly high-risk patients, all men with uncomplicated SUI (no previous surgeries and no history of irradiation) were cured after device implantation. There is evidence, that prior SUI surgeries and radiotherapy might have a negative impact on success rates. These risk factors must be considered for interpretation of the current results.
Concluding message
Despite the above average number of patients with risk factors for failure in this cohort, the results of the current trial are consistent with the evidence in literature and confirm the effectiveness of this new adjustable AUS for treatment of male SUI.
References
  1. Ziegelmann MJ, Linder BJ, Rivera ME, Viers BR, Elliott DS. The impact of prior urethral sling on artificial urinary sphincter outcomes. Can Urol Assoc J. 2016;10(11-12):405- 9.
  2. Cohen AJ, Kuchta K, Park S, Milose J. Patterns and timing of artificial urinary sphincter failure. World J Urol. 2018;36(6):939-45.
  3. Crivellaro S, Morlacco A, Bodo G, Agro EF, Gozzi C, Pistolesi D, et al. Systematic review of surgical treatment of post radical prostatectomy stress urinary incontinence. Neurourol Urodyn. 2016;35(8):875-81.
Disclosures
Funding None Clinical Trial Yes Public Registry No RCT No Subjects Human
01/07/2024 03:23:10